Evaluation of the shielding in a treatment room with an electronic brachytherapy unit.
2017
Esteya® (Elekta Brachytherapy, Veenendaal, The Netherlands) is an electronic brachytherapy (eBT) system based on a 69.5 kVp x-ray source and a set of collimators of 1 to 3 cm in diameter, used for treating non-melanoma skin cancer lesions. This study aims to estimate room shielding requirements for this unit. The non-primary (scattered and leakage) ambient dose equivalent rates were measured with a Berthold LB-133 monitor (Berthold Technologies, Bad Wildbad, Germany). The latter ranges from 17 mSv h-1 at 0.25 m distance from the x-ray source to 0.1 mSv h-1 at 2.5 m. The necessary room shielding was then estimated following US and some European guidelines. The room shielding for all barriers…